Technical characteristics
100,000 cts digital multimeters 57
Technical characteristics of the MTX 3292, MTX 3293 (continued)
Specifications valid from 20% of the range in A, 10% of the range in V
The peak factor is calculated: CF = (Pk+ - Pk-)/2xVrms
Additional error for 250µs < t < 500µs: 3%
Remark: measurements time-stamped only on the MTX 3292, MTX 3293
Accuracy and rate: same as Volt and Ampere measurement specifications
Display of the measurement in dBm with respect to a resistance reference chosen by the
user between 1 and 10k (default value 600).
Resolution 0.01dBm
Absolute error in dBm 0.09 x relative err. VAC expressed in %
Additional calculation error 0.01dBm
Measurement span 10mV to 1,000V
Protection 1,414 Vpk
Display of the measurement in dB with the value measured when the mode was activated
as voltage reference (V ref.).
Resolution 0.01dB
Absolute error in dB 0.09 x relative err. VAC expressed in %
Additional calculation error 0.01dB
Measurement span 10mV to 1,000V
Protection 1,414 Vpk
Display of the measurement in relative power with respect to a resistance reference chosen
by the user between 1 and 10k (default value 50).
The function determined is: (measured voltage)
2
/W Ref (W unit)
(measured current)
2
W Ref (W unit)
Range DC, AC and AC+DC
Resolution 100 W
Accuracy 2 x accuracy in VDC/VAC expressed in %
Max. measurement voltage 1,000 VAC + DC
Protection 1,414 Vpk
Unit of display W
In AC and AC+DC voltage measurement: this calculation is limited to 400Hz.
The current measurement is always made in AC+DC.
Accuracy (typical)/Accuracy of V measurement + Accuracy of peak A measurement
The link to the COM input must be short and of large diameter, in order to limit
the voltage drop that influences the Volt measurement.
Protection: 1,414 Vpk