Technical characteristics
56 100,000 cts digital multimeters
Technical characteristics of the MTX 3292, MTX 3293 (continued)
The user can measure the temperature by means of a Pt100/Pt1000 sensor.
<1mA (Pt100)
<0.1mA (Pt1000)
<1mA (Pt100)
<0.1mA (Pt1000)
0.1% L 1°C
0.07% L 1°C
“Active” protection by PTC thermistor
Display in °C/°F possible
Specified measurement domain
-10.0°C to + 60.0°C
+ 14.0°F to + 140.0°F
-40.0°C to + 999.9°C
-40.0°F to + 1,831.8°F
+ 1,000°C to + 1,200°C
+ 1,832°F to + 2,192°F
Thermal time constant (note 2)
Depending on model of sensor
Detection of sensor open-circuit
Yes: indication of the internal temperature
although the external sensor is connected
Note 1: The stated accuracy in external temperature measurement does not take into
account the accuracy of the K thermocouple.
Note 2: Operation of the thermal time constant (0.7 min/°C):
If there is a sudden variation of the temperature of the multimeter, by 10°C for
example, the multimeter will reach 99% of the final temperature at the end of
5 time constants, or 0.7min/°Cx10°Cx5 cts = 35 min (to which must be added the
constant of the external sensor)
Protection: 1,414 Vpk
www.ShopAEMC.ca
Shop for AEMC products online at:
1.800.561.8187