101
Remove old fuses from holder and discard.
Install the new fuses as shown.
Fuse Ratings:
5A 250 VAC, 5 x 20mm
Slo-Blo
(2) Required
They clip in to the plastic holder, and should
be positioned at the center of the holder.
Close the fuse door fully. Plug the AC power
cord in. (Plug the RJ-45 cable in if it was
removed.)
Turn on AC power and re-establish
instrument communication.
8.5 Maintenance Log
Vision provides a Maintenance Log in the Diagnostic Database to permit tracking of maintenance
activity. This provides a convenient place to find information about tests, lamp changes, and user-
entered comments.
To access the Maintenance
Log, click on Diagnostics,
Maintenance Log, then the
correct selection.