69
7.1.4 Instrument Calibration
Instrument Calibration uses an NIST-
traceable, stable, standard, of known
wavelength response, as a method to
establish wavelength scale response of
the instrument.
The instrument is set to scan the
standard, and the nominal peak
positions for each major absorption are
determined. Vision performs an
algorithm to set the peak positions of
the instrument to those of the standard.
These adjustments are saved, and are
applied on each subsequent scan of the
instrument, yielding a correct spectrum.
Vision takes an instrument reference, which takes
about 20 seconds. This dialog box is displayed.
Vision determines the detector “time constants” using
a patented algorithm, and sets the proper correction
for this parameter.
Wavelength Linearization is performed in two
sections, each of which takes about 45 seconds.
No user response is required during this test.
Another reference scan is taken, and this box is
displayed. This takes about 20-30 seconds.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Metrohm FOSS NIRS XDS MultiVial Analyzer and is the answer not in the manual?

Metrohm FOSS NIRS XDS MultiVial Analyzer Specifications

General IconGeneral
BrandMetrohm
ModelFOSS NIRS XDS MultiVial Analyzer
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals