EasyManua.ls Logo

Metrohm FOSS NIRS XDS MultiVial Analyzer - Sampling Mechanism Not Moving with Large Sample Cell; X-Y Mechanism Not Moving with Vial Tray; X-Y Mechanism Crashing into Iris Adapter; Instrument Cooling Fans Not Operating

Metrohm FOSS NIRS XDS MultiVial Analyzer
118 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
112
▪▪▪▪▪▪▪
CARRIAGE
Error message. Disconnect from the instrument,
cycle instrument power, then re-connect. This will
clear the X-Y error.
IMPORTANT NOTE: The operator should never
attempt to manually position, push, or otherwise
force the mechanism to any position. This can
cause the assembly to “forget its position.” A reset
is required if this occurs.
10 Sampling Mechanism does not
move when using large
sample cell
Click “Use Boat in Vial Dialog box.
11
X-Y Sampling Mechanism does
not move when using vial tray
Click “Use Tray” in Select Tray Dialog box.
12
X-Y Sampling Mechanism
crashes into Iris Adapter,
causing Limit Switch Error
Click “Use Iris” in Vial Dialog box.
13
Instrument Cooling Fans do
not operate
1. Instrument may not be warmed up. Fans do not
operate until the instrument is near operating
temperature.
2. Ambient temperature may not require fan cooling.
This is common in cool environments.
To check fan operation, cycle the Power Switch.
Note if the fans come on for a short burst,
approximately 25 seconds after power, the fans
are operating properly.
If these measures do not correct the problem, please contact Metrohm service.

Table of Contents

Related product manuals