EasyManua.ls Logo

Metrohm FOSS NIRS XDS MultiVial Analyzer - Page 26

Metrohm FOSS NIRS XDS MultiVial Analyzer
118 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
24
▪▪▪▪▪▪▪
DO NOT OPERATE OR TROUBLESHOOT WITH THE LID OPEN. SUNLIGHT MAY CONFUSE
THE OPTICAL INTERRUPTERS USED TO POSITION THE SAMPLE CARRIAGE. THIS MAY
CAUSE AN “X-Y POSITIONING ERROR” TO BE REPORTED.
Follow the assembly sequence that begins on the next page:
1
Load Vision Spectral Analysis Software onto
the computer designated to operate the
XDS instrument.
2
Place the monochromator on the lab bench
in the position shown.
3
Open the right-hand panel of the
instrument. Pull it gently by a fin, until the
catch releases. This panel opens to about a
45-degree angle for access to connectors,
and for filter inspection. Avoid scratches or
damage.

Table of Contents

Related product manuals