95
7. With the lamp wires straight, lift the
lamp retaining ring upward and off the
lamp as shown.
Discard this lamp immediately. Order a
new lamp from your Metrohm
distributor to keep for the next lamp
change.
Always keep a spare lamp in stock, to
avoid last-minute emergencies. The part
number is XA-3000.
8. Notice the black arrow on the new
lamp. This arrow must be upward when
the lamp is installed into the instrument.
This will be shown later.
9.
Unpack the new lamp from the
protective box. Place the new lamp
carefully onto the bench, and place the
lamp retaining ring over the wires as
shown.