1. Press Equipment Classification (F6) and select classifica-
tion.
2. Press MORE (F1).
3. Press Test according to (F7) and select test standard.
4. Press Test Type (F6) and select either Rapid or Normal.
5. Press Test Mode (F5) and select either Automatic or Ma-
nual.
6. Press PL on the keypad to obtain the recording window for
patient lead.
7. Press Module Code (F7) if you wish to record the name of a
new module.
8. Press ENTER (.↵).
9. Press No of leads (F6) to enter the number of leads.
10. Press ENTER (.↵).
11. Press Type (F5) to select protective class.
12. Press ADD (F1) to accept. The number of modules stored
will be shown in parenthesis in the Module Code line.
13. Repeat, as required, to enter a new patient module.
14. Press GO BACK (F2) to return to the menu.
15. Press START (F4) in the MAIN MENU to start the test.
Equipment Code : ...> F7
Sequence Name : ...> F6
Class: CL1 ,0 leads in 0 modules.......> F5
MORE MEMORY SETUP START
F1 F2 F3 F4