• Press ADD (F1) to save in the memory. The number of mod-
ules stored will be shown in parenthesis in the Module Code
line.
• Repeat the above guidelines to enter the next module.
• Press NEXT (F4) to go to the next patient module stored.
• Press PREV (F3) to go to the previous patient module stored.
• Press GO BACK (F2) to return to the previous menu.
7. START (F4).
This function starts the test sequence. Manual or automatic test
sequences are selected under MORE (F1) in MAIN MENU.
3.6 Measurements with
Several Modules in
Manual Mode
When you perform measurements on several modules in Manual
Mode and want to select which module to measure, use the following
procedure:
1. Press MORE (F1) in MAIN MENU.
Equipment Code : ...> F7
Sequence Name : ...> F6
Class: CL1 ,0 leads in 0 modules.......> F5
MORE MEMORY SETUP START
F1 F2 F3 F4
2. Press Test Mode (F5) and select Manual. Press START (F4).
Test according to : IEC60601.1 ......> F7
Test Type : Rapid ...........> F6
Test Mode : Manual ..........> F5
GO BACK START
F1 F2 F3 F4
3. Press MORE (F1) in repeatedly until you find the desired mea-
surement, e.g., Mains on applied part (F5).
MANUAL TEST SETUP F7
Mains Voltage .........................> F6