4-8
2. Press Test Mode (F5) and select Manual.
3. Press START (F4).
4. Press MORE (F1) in MANUAL TEST SETUP.
5. Press MORE (F1) three more times.
6. Press Current Measurement Dual Lead (F7).
7. Press START (F1).
Equipment Code : ...> F7
Sequence Name : ...> F6
Class: CL1 ,0 leads in 0 modules.......> F5
MORE MEMORY SETUP START
F1 F2 F3 F4
Test according to : IEC60601.1 ......> F7
Test Type : Rapid ...........> F6
Test Mode : Automatic .......> F5
GO BACK START
F1 F2 F3 F4
MANUAL TEST SETUP F7
Mains Voltage .........................> F6
Current Consumption ...................> F5
MORE MAIN MENU
F1 F2 F3 F4
Current Measurement Dual Lead .........> F7
Voltage Measurement Dual Lead .........> F6
Resistance Measurement Dual Lead ......> F5
GO BACK START
F1 F2 F3 F4
Open Supply ...........................> F7
Result : µA F6
Measure between E+ and E- inputs F5
START GO BACK PRINT MAIN MENU