If the equipment under test is to be tested for voltages and/or frequen-
cies that differs from the nominal mains supply, the test voltage must
be connected to the auxiliary inlet. To route the auxiliary power to the
contact on the front panel, the procedure below must be executed.
1. Press SETUP (F3) in MAIN MENU.
2. Press MORE (F1) in SYSTEM SETUP.
3. Press MORE (F1) three more times.
4. Press External Isolating Transformer (F7) and select Y
(Yes).
Equipment Code : ...> F7
Sequence Name : ...> F6
Class: CL1 ,0 leads in 0 modules.......> F5
MORE MEMORY SETUP START
F1 F2 F3 F4
SYSTEM SETUP F7
Operator : ...> F6
Establishment : ...> F5
MORE STORE CAL. MAIN MENU
F1 F2 F3 F4
External Isolating Transformer : Y ...> F7
Acoustic alarm for failed test : Y ...> F6
Auxiliary Power : Y ...> F5
MORE GO BACK CAL. MAIN MENU
F1 F2 F3 F4
NOTE
The QA-90 power cable
must also be connected.