14. Test P.A.C. 2 - Patient Auxiliary Current: S.F.C. Open Supply
Applicable to Classes 1 and 2, Types B, BF and CF.
Measures the current flowing between one of the Applied Parts and all of the others in parallel
with one supply lead interrupted (S1 = open); normal and reversed polarity using S2.
AP switch is used to obtain all AP combinations.
Maximum auxiliary current Types B and BF: 50 µA DC; 500 µA (Range: AC 0.1 Hz up to 1
kHz). For Type CF: 50 µA (Range: DC and AC up to 1 kHz).