SediGraph III 5120 Service Card Cage and PCBs
Mar 2010 2-11
Card Cage and PCBs
The card cage is located inside of the cabinetry and is used to hold the cards and backplane
PCBs in a central location. It has all of the interconnections needed to interface to the
instrument and the operating computer.
The following is a listing of the PCBs used in the 5120 and their main purpose.
512/17701/011, Backplane PCB
The Backplane PCB’s main purpose is to provide a connection point for the three PCBs
resident in the card cage and the signals needed for operation of the instrument. The PCB
features the following:
• MICBus interface for 3-card slot cage w/slot ID encoding.
• Board ID EEPROM with instrument ID
• System Configuration ID
• Enable jumper for I
2
C Write protect signal
• i386 CPU board interface w/1600 pin DIN connector
• I/O & Control Interface w/96 pin DIN connector
• Cell Stepper PCB interface w/96 pin DIN connector
• Interface connectors for remote Fluid Control PCB
• Connector for front panel LED
• Surface mount design, except for thru-hole connectors
Top Cover Removed - Showing Card Cage

Other manuals for Micromeritics SediGraph III 5120

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Micromeritics SediGraph III 5120 and is the answer not in the manual?

Micromeritics SediGraph III 5120 Specifications

General IconGeneral
BrandMicromeritics
ModelSediGraph III 5120
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals