SediGraph III 5120 Service Interlock Circuitry
Mar 2010 2-19
X-ray tube
The X-ray tube (PN 510/57603/00) used in the SediGraph 5120 is the same type as that used
on previous SediGraph models. It has the same potentials set for its use, 13,600 Volts for the
high voltage accelerant and no more than 3 mA for current is set during calibration.
The x-ray adjusting system, method, and set up are the same as previous models. The springs
(PN 004/25702/00) used in the alignment process are the same as the 5100. The system works
with two alignment adjusters, which are threaded in a manner to cause the tube to move up/
down with one of the adjusters while the other adjuster moves the cell side to side.