Operational Verification TriStar II 3020 Scheduled Maintenance Instructions
32 Jun 09
5. Click Replace all. The Open Script Test File dialog is displayed.
6. Select leakall.svt, then click OK.
7. Click Next on the Service Test dialog to start the test.
8. During the test, several messages will be displayed. Follow the instructions for each
message then press
OK. Take note of the amount of time each step of the test will take.
Click
Next; the second view of the Service Test dialog is displayed.
Data will be inserted into the pane as collected.
9. After the test is finished, a dialog stating the test is complete is displayed.
10. Click OK to close the dialog, then click Close on the test dialog to close the test.
11. When you close the test, a report is generated automatically to the screen.