TriStar II 3020 Scheduled Maintenance Instructions Operational Verification
Jun 09 33
12. Click Save as; the Save As dialog is displayed.
13. Accept the default name displayed in the File name field, then click Save to return to the
report window.
14. Click Close to close the report window.
15. Use the data sheet provided on page 33 to record the outgas rates, leak rates, and other
information from the report. Include out-of-tolerance or questionable results. Check to
see if results are within the PASSED range. If not, make adjustments to fix the out-of-
tolerance values and rerun the Leakall test until all values pass.