6–21
MEASUREMENT MODE
:MEAS:FUNC:MAJOR?
First AC function query.
Parameters:
None.
Response:
Returns the measurement type according to this table:
0 Capacitance
1 Inductance.
2 Reactance.
3 Susceptance.
4 Impedance.
5 Admittance.
Example: 4
indicates that the first measurement is impedance (Z).
:MEAS:FUNC:MINOR?
Second AC function query.
Parameters:
None.
Response:
Returns the measurement type according to this table:
0 Q-Factor.
1 D-Factor.
2 Resistance.
3 Conductance.
Example: 1
indicates that the second measurement is dissipation factor (D). Note that if the first
measurement is polar (Z or Y), this query will return the last non-polar setting.