STAT ECG
In the event that a STAT ECG is required, hook up the patient using a STANDARD 12-lead hook-up to the IQecg
®
and click STAT ECG
to immediately access the live ECG test screen without having to enter the patient information. A live ECG report can be printed without
saving the patient test, or you can click Analyze to have the computer analyze the data and save the report. The system will prompt the
user to enter the patient information when exiting the Report Review screen.
WARNING
Follow the standard 12-lead ECG placement when performing a STAT ECG. The
automated ECG analysis algorithm assumes the standard 12-lead ECG placement.
Any deviation from the standard 12-lead ECG placements may affect the accuracy
of the automated interpretation.
The green LED on the ECG module will light when the module is
on.
The laptop computer’s AC adapter may introduce electrical
interference. For best ECG results, do not use the AC adapter
while running a live ECG.