Electromagnetic compatibility
This machine has been tested for electromagnetic compatibility
(EMC) in accordance with EN61326-1 and is suitable for operation in
commercial environments, such as laboratories and other similar
environments which are connected to the mains power supply.
The machine's HF emissions are very low and are therefore unlikely to
interfere with other electronic appliances in the vicinity.
Flooring in the installation area must be wood, concrete or tiled.
Synthetic flooring must be able withstand a relative humidity level of
30% to minimise the risk of electrostatic discharges.
The quality of the power supply should comply with that found in a
typical commercial or hospital environment and should deviate from
the nominal voltage by a maximum of +/-10%.