Cleaning the spray arms
The spray arms can become clogged.
Check them daily and clean if needed.
^
Use a sharp pointed object to push
any particles into the spray arm jets,
then rinse thoroughly under running
water.
To do so, remove the spray arms as
follows:
^
Remove the mobile unit.
Washer-disinfector spray arms
^
Loosen the nut with a wrench and
pull the spray arm either upward or
downward.
Spray arms on mobile unit / inserts
^
Use a wrench to loosen the nut on
the spray arm guide, then pull the
spray arm downward.
If you notice wear and tear on the
bearing liners or the gasket, please
contact Miele Customer Service.
^
After cleaning, put the spray arms
back in place and tighten the nut.
Make sure the nut is in the proper
position.
After replacing the spray arms,
check to make sure that they rotate
freely.
,
Make sure that no metallic parts
or pieces become attached to the
magnets at the ends of the spray
arms.
Attached metallic objects can cause
the spray arm speed (RPM) to be
miscalculated.