Important safety instructions
,
Note that these Safety
instructions are in addition to those
found at the beginning of the
Installation and operating
instructions.
~
H
2
O
2
solution must only be used in
special containers from "Ecolab" or "Dr.
Weigert" with the matching adapters.
~
ALWAYS read and comply with the
material safety data sheets from the
manufacturer.
~
Exercise care when handling H
2
O
2
solution - it is a corrosive substance!
Always comply with all applicable
safety regulations!
Use protective eyewear and gloves.
~
Follow the manufacturer's
instructions for disposal of containers.
~
H
2
O
2
solution must NEVER be mixed
with any other processing chemicals.
There is the danger of a violent
chemical reaction, e. g., fulmination
(explosion).
~
Only special tenside detergents
from "Ecolab" or "Dr. Weigert" should be
used.
Connecting the H
2
O
2
container(s)
The line for the H
2
O
2
reservoir is tagged
with a black label. It is delivered without
an adapter because the extraction
systems for each reservoir differ by
manufacturer.
^
Connect the appropriate adapter
from the manufacturer to the (black)
connection line.
^
Connect the H
2
O
2
reservoir(s)
^
Start the service program
DOS2-FILL.
In contrast to the other processing
chemicals, the H
2
O
2
reservoir should
be allowed to completely empty before
you exchange it.
Wait until you see
Fill DOS 2 container
in the display before installing a new
container with H
2
O
2
then start the ser-
vice program
DOS2-FILL.
If the display shows
Check dispensing
system 2
, check the reservoir and
dispensing line. The program
automatically pauses for this.
Special options - OXIVARIO and ORTHOVARIO
60