Height 250 cm (minimum room height)
Width 115 cm
Depth 87 cm
Weight (net) approx. 550 kg
Operating weight approx. 800 kg
Voltage (volts) see data plate
Connected load (watts) see data plate
Fuse rating see data plate
Air pressure connection 600 kPa (required for steam operation)
Steam connection:
with Electric TA
with Steam TA
250 - 1000 kPa / 140 - 180 °C
250 - 600 kPa * / 140 - 164 °C
Water pressure (flow rate) 200 - 1000 kPa overpressure
Cold, hot, and
deionized (DI) water
up to max. 70 °C
Room temperature 5 °C to 40 °C
Relative humidity:
maximum
linear decreasing to
80 % for temperatures up to 31 °C
50 % for temperatures up to 40 °C
Altitude above sea level up to 1500 m
#
Contamination level P2 (as per IEC/EN 61010-1)
Ingress protection rating (as per IEC
60529)
IP20 (ingress of dust)
Noise level in dB (A),
peak LpA for cleaning and drying
cycles
<70
Certification marks CSA, Medical Class II Device Licence
!mark MPG Guidelines - 93/42/EEC, Class IIa
* At a steam pressure of 250 - 300 kPa, the drying unit reaches a maximum drying
temperature of 90 - 100°C.
#
If the machine is installed above 1500 m, the wash water will boil at a lower
temperature. This may require you to lower the disinfection temperature and
lengthen the holding time (A
0
value).
Technical data