Ophthalmology
The E 929/1 mobile injector unit is
designed for the cleaning and
disinfecting of opthalmic instruments.
The final rinse should be carried out
using de-ionised water with a
conductivity of ~15 ZS/cm
(Microsiemens per centimetre). In terms
of quality, the water used for the final
rinse must contain minimal amounts of
endotoxins and pyrogens.
The top level is fitted with various
connections for hollow instruments, e.g.
rinsing and suction hand pieces, and
cannulae.
Instruments are secured to the hose
connectors in the mobile injector unit
using the silicone holders and
connectors in the rack.
The lower level of the mobile injector
unit is designed to take E 441/1 inserts
or E 142 mesh trays for reprocessing
solid instruments.
,
Plastic fibre cover nets must not
be used in washer-disinfectors
which are used to reprocess hollow
opthalmic instruments with narrow
lumens.
The mobile injector unit for
ophthalmic instruments is supplied
with its own operating instructions.
,
The E 929/1 mobile injector unit
for opthalmic instruments may only
be used in washer-disinfectors
which do not have a built-in
dispensing system with a flow rate of
465 ml/min.
,
Washer-disinfectors fitted with
DK+WRG must not be used for
reprocessing opthalmic instruments.
Areas of application
31