This washer-disinfector can be set up
to document processes used.
This can be recorded by either using
external software for process
documentation or an external printer.
Network connection
A 5 m interface cable is supplied for
connecting the washer-disinfector to
suitable network components (e.g. hub,
switch). It is connected to the
Profitronic ethernet interface and is
coiled up in the control cabinet of the
washer-disinfector.
The cable may only be connected to
another interface by a Miele
approved technician.
Serial interface
The cable supplied can also be used to
connect the washer-disinfector to an
external appliance via the Profitronic
RS 232 interface. An RJ45
adapter/ Sub D plug is also supplied.
The interface configuration is
described in section "PC/Printer
functions" of the Programming
manual.
Please note the following when
connecting a printer or PC:
–
Only use an industry-standard PC or
printer (EN/IEC 60950).
–
The type of the printer or PC must be
taken into account when installing
the machine.
–
The maximum length for an extension
cable from the serial interface to the
printer or PC is 10 m and from the
ethernet interface 100 m.
–
A cross-over cable is required for
direct connection between the
washer-disinfector and a PC.
Please contact Miele if you require
further information about suitable
printers and software.
Process documentation
63