Process validation
The standard of cleaning and
disinfection in the disinfection
programmes must be confirmed by the
user as a routine matter.
Safety checks must be carried out to
the machine every 3 months by a Miele
approved service technician in
accordance with EN ISO 15883 and
HTM 2030.
Routine checks
Before the start of each working day the
user must carry out a number of routine
checks. A check list is supplied with the
machine for this purpose.
The following need to be inspected:
All filters in the wash cabinet
The spray arms in the machine and
in any mobile units or modules
The wash cabinet and the door seal
Mobile units, modules and inserts
Maintenance
65