Periodic checks
This washer-disinfector should be
inspected in accordance with local and
national safety regulations after every
2000 operating hours, or annually by
a Miele approved service technician.
This maintenance will cover the
following:
Electrical safety tests
Door mechanism and door seal
Any screw connections and
connectors in the wash cabinet
Water inlet and drainage
Internal and external dispensing
systems
Spray arms
Filter combination
Sump including drain pump and
non-return valve
Steam condenser
All mobile units, modules and inserts
and where applicable:
The drying unit
Any printer connected to the
machine
The following operational tests will be
carried out within the framework of the
maintenance:
A programme will be run as a test
run
Thermo electrical measurements will
be taken
Seals will be tested for water
tightness
All relevant measuring systems will
be safety tested including error
message displays
All safety mechanisms.