Miele PG 8528 - Connecting the H2 O2 solution container

Miele PG 8528
84 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Connecting the H
2
O
2
solution
container
The connection hose for the H
2
O
2
solution container is identified by a
black label. The connection hose is
supplied without a safety cap as the
container for hydrogen peroxide will
vary depending on supplier.
^
Connect the supplier's safety cap to
the connection hose (black).
^
Then connect this to the H
2
O
2
solution container.
^ Then start the
DOS2-FILL
programme.
Unlike with other chemical agents the
H
2
O
2
solution container must be fully
emptied before it is exchanged for a
new one.
^ Do not replace the H
2
O
2
solution
container with a new one until the
Fill
DOS2 container
message is
displayed. After connecting the new
container, you must run the
DOS2-FILL programme.
^
If the
Check dispensing system 2
message appears you should check
the container as well as the
dispensing system. The programme
will stop automatically.