Mobile units, modules and
inserts
Mobile units, modules and inserts
should be checked daily to make sure
they are functioning correctly. The
washer-disinfector is supplied with a
check list.
The following need to be inspected:
Check that the guide rails on mobile
units/inserts are free of hindrance
and that they are secured to the
mobile units/inserts correctly.
Check that the counternuts on the
guide rails are tightened correctly.
Check that the mobile unit connector
is at the correct height and screwed
on correctly.
Check that the locking caps in the
module connectors of modular
system mobile units are working
properly.
Check that jets, sleeves and hose
adapters are securely held in
position in mobile units/inserts.
Check that washing solution can flow
unhindered through all jets, sleeves
and hose adapters.
Make sure caps and closures on
sleeves are correctly located.
Make sure that there are no small
metallic objects sticking to the
magnetic strip on a mobile unit with
automatic mobile unit recognition.
Check that the magnetic strip on
mobile units with automatic mobile
unit recognition is correctly screwed
into position.
Where there are spray arms:
Make sure that the spray arms rotate
freely.
Make sure the spray arm jets are free
of any blockages. See "Cleaning the
spray arms".
Check that there are magnets on the
ends of the spray arms.
Make sure no metal parts have been
attracted to the magnets on the ends
of the spray arms.
In line with the periodic maintenance
of the washer-disinfector the mobile
units, modules and inserts should be
routinely maintained every 2000
operating hours or annually by a
Miele approved service technician.
See ""Maintenance / Periodic
checks".