EasyManua.ls Logo

Miele PG 8593 - Electromagnetic Compatibility (EMC)

Miele PG 8593
184 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Installation
*INSTALLATION*
156
Electromagnetic compatibility (EMC)
The machine has been tested for electromagnetic compatibility (EMC)
in accordance with EN61326-1 and is suitable for operation in
commercial environments, such as hospitals, medical practices and
laboratories and other similar environments which are connected to
the mains power supply.
The machine‘s high frequency (HF) energy emissions are very low and
are therefore unlikely to interfere with other electronic machines in the
vicinity.
Flooring in the installation area must be wood, concrete or tiled.
Synthetic flooring must be able withstand a relative humidity level of
30% to minimise the risk of electrostatic discharges.
The quality of the power supply should comply with that found in a
typical commercial or hospital environment. Check that the power
supply voltage is within a range of +/-10% of its nominal value.