HistosMATE
Automatic cleaner
Operator Manual - MM063

Questions and Answers

  • K
    Kenneth BuckleyAug 9, 2025
    What to do if paraffin traces are still present in Milestone Air Cleaner after cleaning?
    • J
      jariasAug 9, 2025
      If paraffin traces remain after cleaning cycles, and you've replaced the reagents correctly, ensure the mechanical stirrer is moving without obstruction. Also, check if the rinsing ethanol is heavily contaminated with paraffin.