9. ALLARMS- How to manage an alarm
9.1. How to manage an alarm
Serial communication fault
(16 sec)
1. Press ‘Silent’ to hide the alarm.
2. Reboot the system. (wait 2 minutes before the rebooting).
3. Call your customer service if the issue can’t be fixed.
Stirrer not running (20 sec)
1.Check that in the cavity there is nothing that interfere with the
proper stirrer rotation
2.Press ‘Silent’ to hide the alarm.
3.Reboot the system. (wait 2 minutes before the rebooting)
4.Call your customer service if the issue can’t be fixed.
Temperature too high
(5 sec above threshold)
1.Check that reagent level is correct
2.Press ‘Silent’ to hide the alarm.
3.Reboot the system. (wait 2 minutes before the rebooting)
4.Call your customer service if the issue can’t be fixed.
Temperature too low
(5 sec below threshold)
1.Check that reagent level is correct and that a cold reagent
has not been inserted into the cavity.
2.Press ‘Silent’ to hide the alarm.
3.Reboot the system. (wait 2 minutes before the rebooting)
4.Call your customer service if the issue can’t be fixed.