9. ALLARMS- How to manage an alarm
9.1. How to manage an alarm
CODE
DESCRIPTION
WHAT TO DO
1
Serial communication fault
(16 sec)
1. Press ‘Silent’ to hide the alarm.
2. Reboot the system. (wait 2 minutes before the rebooting).
3. Call your customer service if the issue can’t be fixed.
2
Stirrer not running (20 sec)
1.Check that in the cavity there is nothing that interfere with the
proper stirrer rotation
2.Press ‘Silent’ to hide the alarm.
3.Reboot the system. (wait 2 minutes before the rebooting)
4.Call your customer service if the issue can’t be fixed.
3
Temperature too high
(5 sec above threshold)
1.Check that reagent level is correct
2.Press ‘Silent’ to hide the alarm.
3.Reboot the system. (wait 2 minutes before the rebooting)
4.Call your customer service if the issue can’t be fixed.
4
Temperature too low
(5 sec below threshold)
1.Check that reagent level is correct and that a cold reagent
has not been inserted into the cavity.
2.Press ‘Silent’ to hide the alarm.
3.Reboot the system. (wait 2 minutes before the rebooting)
4.Call your customer service if the issue can’t be fixed.