Weight Specifications
The location where the Water Purification System is installed will need to fully support its
Operating Weight:
System type Dry weight (Kg/Lb) Shipping weight (Kg/Lb)
Milli-Q
®
HX 7040 78 / 172 97 / 213
Milli-Q
®
HX 7080 86 / 190 105 / 231
Milli-Q
®
HX 7120 94 / 208 113 / 249
Milli-Q
®
HX 7150 105 / 232 124 / 273
Milli-Q
®
HR 7060 72 / 166 91 / 207
Milli-Q
®
HR 7120 75 / 179 94 / 220
Milli-Q
®
HR 7170 78 / 192 97 / 233
Milli-Q
®
HR 7220 84 / 221 103 / 252
Shipping Weight is defined as a dry system in its shipping container. It does not include
the consumable or any accessories.
Dry Weight is defined as a system without its shipping container. It does not include the
consumable or any accessories.
The UF pre filtration option weight is 22 kg / 121.3 lb
Electrical Specifications
The systems can be supplied by Mains electricity 90-253 VAC with a frequency range from
48 to 62Hz.
System type Voltage Power Consumption (VA)
Milli-Q
®
HX 7040/7080 220-240 VAC @ 50/60 Hz 750
Milli-Q
®
HX 7120/7150 220-240 VAC @ 50/60 Hz 870
Milli-Q
®
HX 7040/7080 120 VAC @ 60 Hz 775
Milli-Q
®
HX 7120/7150 120 VAC @ 60 Hz 900
Milli-Q
®
HX 7040/7080 100 VAC @ 50/60 Hz 775
Milli-Q
®
HX 7120/7150 100 VAC @ 50/60 Hz 900
Milli-Q
®
HR 7060/7120 220-240 VAC @ 50/60 Hz 620
Milli-Q
®
HR 7170/7220 220-240 VAC @ 50/60 Hz 750
Milli-Q
®
HR 7060/7120 120 VAC @ 60 Hz 620
Milli-Q
®
HR 7170/7220 120 VAC @ 60 Hz 750
Milli-Q
®
HR 7060/7120 100 VAC @ 50/60 Hz 620
Milli-Q
®
HR7170/7220 100 VAC @ 50/60 Hz 750
The source of electrical power must be earth grounded.
Specifications | 13