Dispose of old pack.
Remove caps on new pack.
Moisten 0-rings.
Maintain Milli-Q Plus in position with left hand. Slide pack into
position with right hand.
Relock pack retaining arm.
Press OPERATE/STANDBY.
Refer to Section C, Operation, for sequence of steps from this
point,
Normal Operation.
Pack Replacement Frequency
See Section D., Maintenance, 5., Pack Replacement Frequency
Setting.
9. Final Filter (Millipak) Replacement
(Refer also to Section B., Installation, 7., Initial Start-Up)
The final filter unit should be replaced when the flow rate
decreases to an unacceptable level (0.5
1/min).
Press OPERATE/STANDBY to switch the system to STANDBY mode.
Press on tabs at side of point-of-use cover. Remove point-of-use
cover and lay aside.
Unscrew Millipak from final 1/4" female fitting and dispose of
Millipak.
Wrap Teflon tape once around the threads on the 1/4" NPTM end
fitting on Millipak.
Screw Millipak into 1/4" female fitting.
Press OPERATE/STANDBY to switch the system to operate mode.
Loosen vent cap on top of Millipak final filter unit (to vent
air from