Page 16 of 32
4.7 Pump internal flow
sanitization
Materials:
1. Sanitization kit (item MMSANKIT1)
2. Consumables for sanitization kit
(item MMSANSYFU)
3. Sanitizer (see table)
4. Purified water
Remove the pump head
Place a sanitization kit on one pump head
support and lock it
Lock the sanitization kit plug on the other
support
Optional: for faster sanitization, it is
possible to replace the plug by a second
sanitization kit and sanitize the complete
pump at once.
Rotate the trigger counter clockwise, the
switch turns green
Figure 33 trigger is rotated counter clockwise
Take one sanitization funnel assembly
(sanitization kit consumable) and remove
its support, then place the funnel on the
kit
Take a syringe (sanitization kit
consumable) and remove its plunger
Place and lock the syringe on the kit
Pour 60 mL of sanitizer in the syringe
Introduce the plunger into the syringe
Note: An air pocket is visible over the
liquid.
Push the plunger to bring the liquid level
to the 50 mL level indicator
Figure 35: Plunger pushed to bring the liquid to
the 50 mL level indication of the sanitization
syringe
Open the funnel lid
Pour purified water up to the 250 mL
indication level
Put back the funnel lid
Touch the switch to start the pump