7. Culture media growth promotion test
The culture media growth promotion
test, as described in the fully harmonized
pharmacopoeias (USP <61>, EP 2.6.12,
JP 4.05 part 1), can be performed using
the growth promotion accessory (item
MMRECVERY), which is a Milliflex Oasis
®
funnel without membrane.
This accessory allows for good inoculum
spreading and includes a lid to close the
media plate.
This section describes the use of this
accessory:
Obtain and prepare the appropriate
microbiological cultures according to
internal SOPs or pharmacopeia guidelines.
Concentrations should be less than or
equal to 100 Colony Forming Units (CFU)
per sample
Place the media plates in the working area
Place a tray of Milliflex Oasis
®
funnels
without membranes (item MMRECVERY) in
the working area
Remove the media plate cover
Remove the funnel assembly support
Place the funnel assembly on the
media plate
Make sure the funnel assembly is firmly
pressed down against the media plate
Remove the funnel by pinching it while
maintaining the base with the other hand
Tilt the funnel lid sideways to remove it
and store it under the workspace. Discard
the unused funnel
Dispense the inoculum sample onto the
surface of the media plate assembly
and use a sterile spreader to spread the
inoculum on the medium
Note: it is recommended to spread no more
than 50 µL on a media plate assembly. For
method consistency, the same volume must
be spread on the control Petri plate.
Cover the spiked media plate assembly
with the funnel lid
Incubate the media plate assembly at the
specified temperature and time
Note: Incubate the media plate assembly
with the lid facing down.