Milliflex
®
empty cassette for liquid
media (item MXLMC0120)
Use the following steps to prepare a liquid
medium cassette:
Remove the yellow cap from the
cassette
1
Open the ampoule of medium by
twisting off the ampoule top.
2
Insert the male Luer tip of the ampoule
into the female Luer opening of the
cassette.
Squeeze the ampoule to dispense the
medium into the cassette. Use the entire
content of the ampule.
3
Note: Avoid pressing the cassette on
the working surface while inserting the
media, since this might compress the pad
and slow the introduction of the media.
Remove the empty ampoule and firmly
attach the yellow cap. The protective
Tyvek
®
paper cover remains on the
cassette. Remove it before using the
cassette.
Yellow
Cap
Liquid Medium
Cassette
1
Liquid Medium
Ampoule
Ampoule
Twist-o Top
Squeeze the
tube until empty
Tyvek
®
Paper Cover
2 3
Figure 49: steps to prepare a liquid media cassette