EasyManuals Logo

Mindray BA-88A Operation Manual

Mindray BA-88A
107 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #36 background imageLoading...
Page #36 background image
3-4
Endpoint
1 Prepare the samples and calibrators and controls.
Prepare proper amount of reagent according the number of the samples.
If it is double-reagent analysis, it is recommended that R1 and R2 are
mixed and then dispensed into respective tubes (or other reaction
containers).
2
NOTE
Check the reagent instruction, or consult the
reagent manufacturer or distributor to see if mixing
of R1 and R2 are allowed prior to testing.
3 Dispense sample or calibrator or control into the tube.
4 Incubate the tube in a water bath tank or external incubator for a while.
5 Take the tube out to prepare it for analysis.
6 Click Test on the main screen to enter the test screen.
7 Select the name of the test to be run.
8 Click Water Blank, put the inlet tube into the distilled water, and click
Aspirate.
9 Click Reagent Blank, put the inlet tube into the reagent blank, and click
Aspirate.
10 Select Calibrate and select calibrator in the drop down list box of S1-S8.
Put the inlet tube into the calibrator to be tested and click Aspirate. If it is
multiple-point calibration, you can proceed to the next step only when the
calibration tests of all the calibrators are completed. Put the inlet tube
into the sample blank, and click Aspirate.
12
NOTE
You can determine whether it is necessary to run
sample blank.
Hemolysis, icterus or lipemia in the samples may
lead to unreliable test results, so a sample blank is
recommended.
13 Select Sample ID and click Sample. Put the inlet tube into the sample
and click Aspirate. Repeat this to test other samples.
14 After calibration is finished. Click QC and select controls from the
drop-down list box C1-C3. Put the inlet tube into the control to be
tested and click Aspirate.
15 After all the tests are completed, put the inlet tube into the distilled and
click Wash.
Fixed-time and Kinetics
1 Prepare the samples, calibrators and controls.
Prepare proper amount of reagent according the number of the samples. If it is
double-reagent analysis, it is recommended that R1 and R2 are mixed and
then dispensed into respective tubes (or other reaction containers).

Other manuals for Mindray BA-88A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BA-88A and is the answer not in the manual?

Mindray BA-88A Specifications

General IconGeneral
Light sourceHalogen lamp
Display7 inch touch screen
PrinterThermal printer
InterfaceRS232
Wavelengths340, 405, 546, 578, 630nm
Carry-over< 1%
Power SupplyAC 100~240V, 50/60Hz
Operating Temperature15-30°C

Related product manuals