EasyManuals Logo

Mindray BA-88A Operation Manual

Mindray BA-88A
107 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #42 background imageLoading...
Page #42 background image
4-4
Parameter Description
NOTE
The delay is mainly to balance the reaction
temperature and eliminate the effect of small
bubbles. For Endpoint and Absorbance analysis, the
delay is usually set as 6-10 sec. For Fixed-time and
Kinetics analysis, it is mainly used to eliminate the
effect of unwanted reactions, usually set as 30-60s.
Refer to the reagent instruction for more details.
Measuring
time
A period for reading the absorbance to be used to calculate the
result. The acceptance range is 1-9999sec (3-9999sec for Kinetics
analysis). The minimum increment or decrement is 1 sec.
NOTE
The measuring time for endpoint analysis is usually
around 5sec; for Fixed-time and Kinetics, please
refer to the reagent instruction and it is usually more
than 30sec.The reading interval is 1 sec for single
wavelength and 2sec for double wavelengths.
Unit
Unit of the result. Select from the drop-down list. The options
include g/L, mmol/L, umol/L, U/L, mg/dL, IU/L, mg/L and void. The
default is void. The units can be added and deleted in Setup.
Precision Precision of the result. Select from the drop-down list 1, 0.1, 0.01,
0.001.The default is 1.
Asp. Vol. If refers to the sample volume (200-9000ul) to be dispensed for the
reaction. The default is 500ul.
Rgt. Blank
Whether to run reagent blank.
NOTE
Reagent blank must be run for endpoint analysis.
For Kinetics, Fixed-time and Absorbance analysis,
refer to the reagent instruction to decide whether to
run.
Temp. The temperature of the cuvette or flow cell. The options include
room temperature, 25, 30 and 37.The default is 37.
Full name Full name of the test. It can be void.
The following table introduces the buttons on the screen.

Other manuals for Mindray BA-88A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BA-88A and is the answer not in the manual?

Mindray BA-88A Specifications

General IconGeneral
Light sourceHalogen lamp
Display7 inch touch screen
PrinterThermal printer
InterfaceRS232
Wavelengths340, 405, 546, 578, 630nm
Carry-over< 1%
Power SupplyAC 100~240V, 50/60Hz
Operating Temperature15-30°C

Related product manuals