4-4
Parameter Description
NOTE
The delay is mainly to balance the reaction
temperature and eliminate the effect of small
bubbles. For Endpoint and Absorbance analysis, the
delay is usually set as 6-10 sec. For Fixed-time and
Kinetics analysis, it is mainly used to eliminate the
effect of unwanted reactions, usually set as 30-60s.
Refer to the reagent instruction for more details.
Measuring
time
A period for reading the absorbance to be used to calculate the
result. The acceptance range is 1-9999sec (3-9999sec for Kinetics
analysis). The minimum increment or decrement is 1 sec.
NOTE
The measuring time for endpoint analysis is usually
around 5sec; for Fixed-time and Kinetics, please
refer to the reagent instruction and it is usually more
than 30sec.The reading interval is 1 sec for single
wavelength and 2sec for double wavelengths.
Unit
Unit of the result. Select from the drop-down list. The options
include g/L, mmol/L, umol/L, U/L, mg/dL, IU/L, mg/L and void. The
default is void. The units can be added and deleted in Setup.
Precision Precision of the result. Select from the drop-down list 1, 0.1, 0.01,
0.001.The default is 1.
Asp. Vol. If refers to the sample volume (200-9000ul) to be dispensed for the
reaction. The default is 500ul.
Rgt. Blank
Whether to run reagent blank.
NOTE
Reagent blank must be run for endpoint analysis.
For Kinetics, Fixed-time and Absorbance analysis,
refer to the reagent instruction to decide whether to
run.
Temp. The temperature of the cuvette or flow cell. The options include
room temperature, 25℃, 30℃ and 37℃.The default is 37℃.
Full name Full name of the test. It can be void.
The following table introduces the buttons on the screen.