Parameter Description
Wash Click Wash. The system will aspirate certain amount of distilled
water to wash the tubing.
Stop Stop testing or washing. The sample ID will not increase
automatically.
Water Blank Run DI water zero.
Rgt. Blank Run reagent blank for the selected test.
Sample Blank Run Sample Blank for the selected sample.
The test result of the next sample will have sample blank
deducted.
For endpoint analysis, you can choose whether to run sample
blank or not based on the status of the sample.
Sample Run selected tests of the current sample. Click Sample. The
following information will be shown in the window above the
reaction curve.
The area shows the sample ID of the current sample. Use “+”
and “-“to change the sample ID. After the completion of a sample
test, the sample ID of the sample blank will increase by 1 (if
requesting is applied, increase the sample ID according to
requesting).
Calibrate Run selected tests of the calibrator. Click Calibrate. The
following information will be shown in the window above the
reaction curve.
The S1-S8 will be displayed in the drop-down list box according
to parameter setup. you can rearrange the testing order of the
calibrators. Unless all the calibration of the calibrators are
completed, system will disable the sample or control test.
QC Run selected tests of the control. Click QC. The following
information will be shown in the window above the reaction
curve.
Select control to run QC test. But the control information should
be preset in the parameter setup.
Water blank When the system running distilled water zero, the ABS of water
blank is displayed
When the system running reagent blank, the ABS of reagent
blank is displayed.

Other manuals for Mindray BA-88A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BA-88A and is the answer not in the manual?

Mindray BA-88A Specifications

General IconGeneral
Light sourceHalogen lamp
Display7 inch touch screen
PrinterThermal printer
InterfaceRS232
Wavelengths340, 405, 546, 578, 630nm
Carry-over< 1%
Power SupplyAC 100~240V, 50/60Hz
Operating Temperature15-30°C