4-22
Buttons  Function 
Print  Print current calibration parameters and curve. 
Return  Click this button to return to the main screen. 
 
CAUTION 
 
The replacement of certain components should be followed with 
calibration. Otherwise, you may get unreliable result. 
 
4.5 QC 
You can view the QC result and QC reaction curve, and check whether the test is out 
of control and print QC reaction curve in the QC screen. 
Figure 4-14 QC 
 
The following table explains the parameters on the screen. 
Parameter  Description 
Test  All the tests with QC information set will be listed. 
Select a test to view the QC result. 
Date  The maximum range is one month(31 days). 
Control  C1-C3 can be selected. 
Lot No.  The latest 5 lot No. will be displayed. The default is the most 
recent lot No..   
SD  Standard Deviation of the control. 
AVG  Concentration of the control.