Mindray BeneFusion SP3 Vet - C Alarm Information; [Time Near End]; [Syringe Near Empty]; [No AC Power]

Mindray BeneFusion SP3 Vet
84 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
C-3
[Time Near End]
Low
Infusion remaining time
reaches the setting value
of [Time Near End].
1. The alarm will not be
cancelled automatically
until the infusion is
completed, and then
switch to [VTBI
Complete] or [Syringe
Empty] alarm.
2. Or press to
cancel alarm.
3. Or press to
acknowledge the alarm
condition.
[Syringe Near
Empty]
Low
Required time for the
remaining liquid within the
syringe reaches the
setting value of [Time
Near End].
1. The alarm will not be
cancelled automatically
until the infusion is
completed, and then
switch to [VTBI
Complete] or [Syringe
Empty] alarm.
2. Or press to
cancel alarm.
3. Or press to
acknowledge the alarm
condition.
[No AC Power]
Low
Power cord disconnects
when the pump has
connected to an AC/DC
power supply.
1. Connect to an AC/DC
power supply to cancel
the alarm.
2. Or press to
cancel alarm.