C-4
[Communication
interrupted]
Low
Syringe pump and
BeneFusion CS5 Infusion
Supervision System are
communicated
successfully over Wi-Fi,
the network
communication is
abnormally interrupted for
3 minutes. After the alarm
is triggered, infusion of the
pump will not be
influenced, and the pump
continues infusion.
Press or restore
the communication
between syringe
pump(s) and
BeneFusion CS5
Infusion Supervision
System.
[KVO Running]
Low
The alarm is triggered
after [VTBI Complete]
alarm is triggered and in
KVO mode.
1. Press to stop
infusion and cancel
alarm.
2. Or the [KVO Finish]
alarm is triggered when
reach the KVO infusion
time.
[Infusion
interrupted]
Low
The alarm is triggered
after any of the
[Occlusion], [Syringe
Disengaged] or [No
Battery] alarm is
cancelled.
Press or to
cancel alarm.
[Please reset time]
Low
RTC time is reset when
the pump is on.
Reset the time or press
to cancel alarm.
[Para.
Unconfirmed]
Low
The alarm is triggered if no
operation is performed
within 10s under the
parameters setting status.
Press any key to cancel
alarm.
NOTE
All alarm sounds can be paused by pressing , except for the
circumstance of [No Battery].