A-10
5. Press SOFT KEYS 1-4 to select leakage tests
6. Press APPLIED PART (SOFT KEY 4) at any time to select the desired applied part
leakage current:
7. Modify the configuration of the front panel outlet by pressing the appropriate SOFT
KEY on the 601PRO:
8. Press the print data key at any time to generate a printout of the latest measurement.
In Case of Failure
Check any broken of the enclosure. Replace any defective part.
Inspect wiring for bad crimps, poor connections, or damage.
Test the wall outlet; verify it is grounded and is free of other wiring abnormalities.
Notify the user or owner to correct any deviations. As a work around, check the other
outlets to see if they could be used instead.
Change another probe to confirm if the fail is caused by console.
If the leakage current measurement tests fail on a new unit and if situation can not be
corrected, submit a Safety Failure Report to document the system problem. Remove unit
from operation.
If all else fails, stop using and inform the Customer Service Engineer for analysis and
disposal.
LIMITS
For CF applied parts,
10μA in Normal Condition
50μA in Single Fault Condition
For BF
applied parts,
100μA in Normal Condition
500μA in Single Fault Condition

Table of Contents

Other manuals for Mindray BeneView T6

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneView T6 and is the answer not in the manual?

Mindray BeneView T6 Specifications

General IconGeneral
BrandMindray
ModelBeneView T6
CategoryMedical Equipment
LanguageEnglish

Related product manuals