3-33
3.4.18 EEG tests
Method 1
Tools required:
EEG simulator (Nannini 169/1)
1. Insert the EEG module into the module slot of the monitor.
2. Connect the EEG module to the EEG simulator through the EEG patient cable.
3. Select [EEG Setup]→ [Sensor Check] to perform a sensor check.
4. Check that the result for each electrode is pass.
5. Check that the EEG wave and numeric are displayed on the screen.
Method 2
Tools required:
None
1. Insert the EEG module into the module slot of the monitor.
2. Connect the EEG cable and EEG electrodes to the EEG module, and short all the
electrodes.
3. Select [EEG Setup]→ [Sensor Check] to perform a sensor check.
4. Check that the result for each electrode is pass.
5. Check that the EEG wave and numeric are displayed on the screen.