3-47
6. Select [Close] to finish.
3.13.2 Restoring the System
Connecting the External USB Cable
Connect the iView maintenance USB disk which stores the system image files and keyboard
to the USB connectors special for iView maintenance.
Setting BIOS Screen
1. Start up the patient monitor and press [F2] key repeatedly on the keyboard.
2. Select [Main Menu] [Maintenance>>] [User Maintenance>>] enter the
required password [iView Maintenance>>] [iView Maintenance], and when
the system switches to BIOS setup screen, stop press [F2] key.
3. Move the cursor to [Boot] by pressing [] or [] keys on the keyboard. Then press
[Enter] on the keyboard.
4. Move the cursor to [1st Boot], and press [Enter] on the keyboard.
5. Select [USB] and press [Enter] on the keyboard.
6. Move the cursor to [Save&Exit], and press [Enter] on the keyboard. Move the cursor to
[Yes] and press [Enter] on the keyboard to save the setup.