P/N: 046-017383-00 Rev1.0
15
BeneVision TM80 Quick Reference Guide
Unpair a TM80 with a BP10 via the TM80
Option 1:
1. Press to enter the main menu.
2. Tap Wireles Modules.
3. Select the device you wish to disconnect from the TM80 from the
devices list under Devices and tap Disconnect.
The connected device is disconnected from the TM80.
Option 2:
1. Press to enter the main menu.
2. Tap Wireles Modules.
3. Tap MPAN.
4. Select Off.
Unpair a TM80 with a BP10 via the BP10
1. Press the MPAN key on the BP10.
2. Select Yes when the prompt message “Are you sure you want to close
MPAN and unpair BP10?” appears.
Caution: Before moving a TM80 or BP10 to another area, unpair them first.
Caution: Before admitting a new patient at a BP10, you need to unpair it with a
TM80 first.
ALARMS
Alarm Status Symbols
Symbol Description
All the alarms are paused.
Audible alarm tones are turned off.
Alarms are acknowledged and the alarm system is reset.