17 Probes and Biopsy
Operator’s Manual 17 - 45
Perform the following procedure:
1. Install the needle-guided bracket:
a. Put on the sterile probe sheath.
b. Align the needle-guided bracket with the probe to align the locating bulges on the needle
guide with the locating grooves on the probe, then turn the retaining clamp to the correct
position.
c. Hold the probe, and insert the needle into Needle entry of the needle guide.
2. Remove the needle-guided bracket:
a. Hold the probe and remove the needle.
b. Hold the probe and raise the needle-guided bracket to separate the locating bulge from the
locating grooves.
3 Needle entry 4 Retaining clamp
5 Locating bulges