17 - 46 Operator’s Manual
17 Probes and Biopsy
NGB-048
Perform the following procedure:
1. Install the needle-guided bracket:
a. Put on the sterile probe sheath.
b. Hold the probe in one hand, select the correct needle-guided bracket and hold it with the
other hand. Press the separating button of the bracket to open the clamp. Match the fixing
structure of the bracket with the positioning tab of the probe. Push the needle guided
bracket forward to insert the clamp into the installation grove at the back of the probe.
Mount the bracket onto the probe.
1 Separating button 2 Needle guiding structure
3Clamp 4Probe fixing structure
1
2
3
4