G - 4 Operator’s Manual
  G  EMC Guidance and Manufacturer’s Declaration
Table  G-3 
GUIDANCE AND MINDRAY DECLARATION-ELECTROMAGNETIC IMMUNITY
Consona N9 series is intended for use in the electromagnetic environment specified below. The 
customer or the user of system should assure that it is used in such an environment.
IMMUNITY TEST IEC 60601
TEST LEVEL
COMPLIANCE 
LEVEL
ELECTROMAGNETIC 
ENVIRONMENT-
GUIDANCE
Electrostatic 
Discharge (ESD)
IEC 61000-4-2
±8 kV contact;
±2 kV, ±4 kV, ±8 kV, 
±15 kV air
±8 kV contact;
±2 kV, ±4 kV, ±8 kV, 
±15 kV air
Floors should be wood, 
concrete or ceramic tile. If 
floors are covered with 
synthetic material, the 
relative humidity should 
be at least 30%.
Electrical fast 
Transient / burst
IEC 61000-4-4
±2 kV for power 
supply lines;
±1 kV for input/output 
lines
±2 kV for power 
supply lines;
±1 kV for input/output 
lines
Mains power quality 
should be that of a typical 
commercial or hospital 
environment.
Surge
IEC 61000-4-5
±0,5 kV, ±1 kV line(s) 
to line(s);
±0,5 kV, ±1 kV, ±2 kV 
line(s) to earth
±0,5 kV, ±1 kV line(s) 
to line(s);
±0,5 kV, ±1 kV, ±2 kV 
line(s) to earth
Mains power quality 
should be that of a typical 
commercial or hospital 
environment.
Voltage dips, Short 
interruptions and 
voltage variation on 
power supply input 
voltage 
IEC 61000-4-11
0% U
T
; 0,5 cycle 
At 0°, 45°, 90°, 135°, 
180°, 225°, 270° and 
315°
0% U
T
; 1 cycle
70% U
T
 for 25/30 
cycle at 0°
0% U
T
; 250/300 cycle
0% U
T
; 0,5 cycle 
At 0°, 45°, 90°, 135°, 
180°, 225°, 270° and 
315°
0% U
T
; 1 cycle
70% U
T
 for 25/30 
cycle at 0°
0% U
T
; 250/300 cycle
Mains power quality 
should be that of a typical 
commercial or hospital 
environment. If you 
require continued 
operation during power 
mains interruptions, it is 
recommended that our 
product be powered from 
an uninterruptible power 
supply or a battery.
Power frequency 
(50/60 HZ) 
magnetic field
IEC 61000-4-8
30 A/m 30 A/m Power frequency 
magnetic fields should be 
at levels characteristic of a 
typical location in a 
typical commercial or 
hospital environment.
NOTE:
U
T
 is the A.C. mains voltage prior to application of the test level.