G  EMC Guidance and Manufacturer’s Declaration
Operator’s Manual G - 5
Table  G-4 
GUIDANCE AND MINDRAY DECLARATION-ELECTROMAGNETIC IMMUNITY
Consona N9 series is intended for use in the electromagnetic environment specified below. The 
customer or the user of system should assure that it is used in such an environment.
IMMUNITY 
TEST
IEC 60601 TEST 
LEVEL
COMPLIANCE 
LEVEL
ELECTROMAGNETIC 
ENVIRONMENT-GUIDANCE
Conduced RF 
IEC 61000-4-6 
3 Vrms 
0,15 MHz – 80 
MHz
6 Vrms in ISM
a
 
and amateur radio 
bands between 
0,15 MHz and 80 
MHz 
3 Vrms 
0,15 MHz – 80 
MHz
6 Vrms in ISM 
and amateur radio 
bands between 
0,15 MHz and 80 
MHz 
Portable and mobile RF 
communications equipment should be 
used no closer to any part of system, 
including cables, than the recommended 
separation distance calculated from the 
equation applicable to the frequency of 
the transmitter. Recommended 
separation distance
d = 1.2 ×
d = 2 ×
Radiated RF 
IEC 61000-4-3
10 V/m
80MHz - 2.7GHz
10 V/m
80MHz - 2.7GHz
d = 1.2 ×  80 MHz to 800 MHz
d = 2.3 ×  800 MHz to 2.7GHz
Where, P is the maximum output power 
rating of the transmitter in watts (W) 
according to the transmitter 
manufacturer and d is the recommended 
separation distance in meters (m).
Field strengths from fixed RF 
transmitters, as determined by an 
electromagnetic site survey
b
, should be 
less than the compliance level in each 
frequency range
c
. 
Interference may occur in the vicinity of 
equipment marked with the following 
symbol: 
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected 
by absorption and reflection from structures, objects and people.